Cart - Calibration

This should be done after your Rated Capacity and Sensitivity Numbers are inputted. ( Click Here for Article )
  1. Ensure load cell ratings (Rated Capacity and Sensitivity) have been found and entered in the Device tab. 
  2. Ensure the Libra Cart device and app are paired and communicating.
  3. Ensure the grain cart is empty and on level ground.
  4. Select the Device tab and tap Zero Scale.
  5. Select the Display tab and fill the grain cart (larger weights provide higher precision). 
  6. Tap Tare and then unload the cart into an empty truck, noting the amount unloaded (record as positive). 
  7. Weigh the load with a trusted scale. 
  8. Select the Device tab and enter the weight from the trusted scale as the Certified Reading. 
  9. Enter the displayed weight unloaded as the Cart Reading.
  10. Tap Calculate to compute a new Calibration Factor. 
  11. Tap Save Settings. 

For additional help the video tutorial below will show this process in-app: