Plant - Calibration
This should be done after your Rated Capacity and Sensitivity Numbers have been entered. Click here for an article on how to set your load cell ratings.
- Ensure the Libra Plant device and app are paired and connected.
- Ensure the equipment is empty and on level ground.
- Select the Device tab and tap Zero Scale on the appropriate weight channel.
- Tap Tare on the Display tab and then apply a known weight to the weight channel's container, noting the amount displayed (record as positive).
- Select the Device tab, tap the desired weight channel's blue Calibration number and enter the known weight as the Certified Reading.
- Enter the noted weight from the app's Display as the Libra Reading.
- Tap Next to compute a new Calibration Factor. Tap Save to confirm Calibration.
- If using both weight channels, repeat steps 3-7 for the other weight channel.